Self-testing, Basic information of the barometer settings – Vaisala PTB210 (serial) User Manual
Page 22

User's Guide ______________________________________________________________________
20 __________________________________________________________________ M210669EN-D
.RON<cr>
The .
RON
command sets the terminating resistor on.
.ROFF<cr>
The .
ROFF
command sets the terminating resistor off.
Self-Testing
.? Basic Information of the
Barometer Settings
.?<cr>
The command lists the basic information of the barometer settings.
The following example shows also the factory settings for 50…1100
hPa model.
.?<cr> (command not visible)
PTB210 Ver 1.0
CAL DATE :1999-06-10
ID CODE :10
SERIAL NUMBER :T00100004
MULTIPOINT CORR:ON
MEAS PER MINUTE: 60
AVERAGING : 0
PRESSURE UNIT : hPa
Pressure Min...Max: 500 1100 (50 1100 for model 50…1100)
LOW CURRENT MODE
RS485 RESISTOR OFF