Chapter 4 ________________________________________________________________ Operating

VAISALA _______________________________________________________________________ 17

Operating Commands

.P Outputting One Single Message

.P<cr>

The command outputs the latest multi-point corrected pressure reading
with two decimals.

.P<cr> (command invisible)
1012.99

.BP Continuous Outputting

.BP<cr>

The command outputs the MPC corrected pressure reading
continuously. The output interval depends on the measurement rate
and the averaging. The outputting is stopped by pressing enter key
(<cr>).

.P<cr> (command invisible)

997.99
998.01
998.01

...
998.02
<cr>

(command invisible)

.UNIT Selecting a Pressure Unit

.UNIT.x<cr>

x = 0 (hPa)

6 (kPa)

1

(mbar)

7

(Pa)

2

(inHg)

8

(mmH

2

O)

3

(psia)

9

(inH

2

O)

4

(torr)

10

(bar)

5

(mmHg

As default, the unit is 0.

NOTE

To activate the new setting, please remember to reset the barometer
with the command .

RESET

<cr>