Chapter 4 ________________________________________________________________ Operating
VAISALA _______________________________________________________________________ 17
Operating Commands
.P Outputting One Single Message
.P<cr>
The command outputs the latest multi-point corrected pressure reading
with two decimals.
.P<cr> (command invisible)
1012.99
.BP Continuous Outputting
.BP<cr>
The command outputs the MPC corrected pressure reading
continuously. The output interval depends on the measurement rate
and the averaging. The outputting is stopped by pressing enter key
(<cr>).
.P<cr> (command invisible)
997.99
998.01
998.01
...
998.02
<cr>
(command invisible)
.UNIT Selecting a Pressure Unit
.UNIT.x<cr>
x = 0 (hPa)
6 (kPa)
1
(mbar)
7
(Pa)
2
(inHg)
8
(mmH
2
O)
3
(psia)
9
(inH
2
O)
4
(torr)
10
(bar)
5
(mmHg
As default, the unit is 0.
NOTE
To activate the new setting, please remember to reset the barometer
with the command .
RESET
<cr>