Chapter 6, Technical data, Technical specification – Vaisala PTB210 (serial) User Manual

Page 27: Operating range, Accuracy, Technical specification operating range

Advertising
background image

Chapter 6 ____________________________________________________________ Technical Data

VAISALA _______________________________________________________________________ 25

CHAPTER 6

TECHNICAL DATA

Technical Specification

Operating Range

Pressure range

500 ... 1100 hPa

50 ... 1100 hPa

Operating temperature range

-40 ... +60

C

Humidity range

non condensing

Accuracy

500…1100 hPa

Class A

ClassB

Non linearity*

 0.10 hPa

 0.15 hPa

Hysteresis*

 0.05 hPa

 0.05 hPa

Repeatability*

 0.05 hPa

 0.05 hPa

Calibration uncertainty**

± 0.07 hPa

± 0.15 hPa

Total accuracy (20

C)***  0.15 hPa

 0.20 hPa

Temperature dependence
(-40 ... +60

C)****

± 0.20 hPa

± 0.20 hPa

Total accuracy***
(-40 ... +60

C)

 0.25 hPa

 0.30 hPa

Long term stability

 0.10 hPa/year  0.10 hPa/year

50…1100 hPa
Non linearity*

 0.20 hPa

Hysteresis*

 0.10 hPa

Repeatability*

 0.10 hPa

Calibration uncertainty**

± 0.20 hPa

Total accuracy (20

C)***

 0.35 hPa

Temperature dependence
(-40 ... +60

C)****

± 0.40 hPa

Total accuracy***
(-40 ... +60

C)  0.50 hPa

Long term stability

± 0.20 hPa/year