Technical summary – Notifier Zareba Sensepoint Pro User Manual

Page 4

Advertising
background image

Technical Summary

General Specification

Use

3 wire, 4-20mA, fixed point gas detector with built in alarm relays for the protection
of personnel and plant from flammable, toxic and oxygen gas hazards. Incorporating
transmitter with local display and fully configurable via remote handheld control

Detectable Gases

Gas

Measuring Range

Operating Temperature*

Ident

Name

1

#

2

3

Min

Max

F

Flammable

0-100% LEL

-20°C / -4°F

50°C / 122°F

O

Oxygen

0-25% VOL

-15°C / 5°F

40°C / 104°F

C

Carbon Monoxide

+

0-200ppm

0-500ppm

-20°C / -4°F

50°C / 122°F

H

Hydrogen Sulfide

+

0-50ppm

0-20ppm

0-100ppm

-20°C / -4°F

50°C / 122°F

G

Hydrogen

0-1000ppm

-5°C / 23°F

40°C / 104°F

+

Surecell

TM

Sensor Technology

#

Standard range

*

Check certified temperature

Electrical

General

3 wire connection, 12-30Vdc, 4-20mA (source), common 0Vdc
7 x screw terminals for O.5mm

2

(20AWG) to 2.5mm

2

(14AWG) wire

Maximum operating current: 300mA (flammable), 150mA (toxic and oxygen)
Maximum peak in rush current: 600mA

Signal

22mA

Maximum over range

4-20mA

Normal operation: 0-100% full scale

2mA or 4mA

Output inhibited (2mA or 17.4mA for oxygen)

0mA

Fault

Alarm Relays

2 x 3A@240Vac alarm relays

User configurable alarm trip point, normally energized/de-energized operation

Link selectable output as NO or NC

Construction

Material

Main body: Powder coated aluminum (ADC12)

Sensor: 316 stainless steel

Weatherproof cap: Aluminum, 316 stainless steel mesh

Weight

1450g

Certification

Sensepoint Pro

Sensepoint Pro Remote Control ZHC1

ATEX

Ex II 2G EEx d IIC T4

Ex II 2G EEx ib IIC T4

(T

amb

= -20°C to +50°C / -4°F to +122°F)

(T

amb

= -20°C to +50°C / -4°F to +122°F)

UL

Class 1, Div 1, Groups B, C, D

Class 1, Div 1, Groups A, B, C, D

(T

amb

= -20°C to +50°C / -4°F to +122°F)

(T

amb

= -20°C to +50°C / -4°F to +122°F)

Applicable Standards

North America: UL1203

Europe: CE, ATEX, EMC (EN50270)

Contact Honeywell Analytics for latest information

Environmental

IP Rating

IP65 as standard in accordance with EN60529:1992

Operating Temperature

Dependent upon gas type and certification

Operating Humidity Range

Continuous: 20-90%RH (non condensing), Intermittent: 10-99%RH (non condensing)

Operating Pressure

90-110kPa

Storage Conditions

Instrument -40°C to +50°C (-40°F to +122°F)

* Oxygen detection products must only be used to detect oxygen depletion in air.