Anritsu MG3641A User Manual
Page 141
SECTION 6 GPIB
6-35
:AMPLitude[ :OUT]:SWEep:STARt?
Function
Reads the start level for a sweep.
Response
Start level for a sweep
Restriction
None
:AMPLitude[ :OUT]:SWEep:STOP <Numeric> <Ampl term>
Function
Sets the stop level for a sweep.
Parameter
<Numeric> = –142.99 dBm to +22.99 dBm
Unit
<Ampl term>
Restriction
None
:AMPLitude[ :OUT]SWEep:STOP?
Function
Reads the stop level for a sweep.
Response
Stop level for sweep
Restriction
None
:AMPLitude[ :OUT]:SWEep:CENTer <Numeric> <Ampl term>
Function
Sets the center level for a sweep.
Parameter
<Numeric> = –142.99 dBm to +22.99 dBm
Unit
<Ampl term>
Restriction
None
:AMPLitude[ :OUT]:SWEep:CENTer?
Function
Reads out the center level for a sweep.
Response
Center level for a sweep
Restriction
None
:AMPLitude[ :OUT]:SWEep:SPAN <Numeric> <Ampl term>
Function
Sets the span level for a sweep.
Parameter
<Numeric> = 0.02 dB to 20 dB
Unit
<Ampl term>
Restriction
Unit is dB, only.
:AMPLitude[ :OUT]:SWEep:SPAN?
Function
Reads out the span level for a sweep.
Response
Span level for a sweep
Restriction
None
:AMPLitude[ :OUT]:SWEep:STEP:SIZE <Numeric> <Ampl term>
Function
Sets the level step size for a sweep.
Parameter
<Numeric> = 0.01 dB to 20 dB
Unit
<Ampl term>
Restriction
None
:AMPLitude[ :OUT]:SWEep:STEP:SIZE?
Function
Reads out the level step size for a sweep.
Response
Step size for a sweep
Restriction
None