Agilent Technologies Switch/Measure User Manual

Page 26

Advertising
background image

14

34980A User’s Guide

2

Features and Functions

Front Panel Operation:

• To configure the measurement parameters and add a channel to the

scan list, use the

Channel (Configure) key.

• To initiate a scan and store all readings in memory, press the

Scan (Measure) key. If you press the Scan (Measure) key with no
scan list defined, the instrument initiates a DMM- only measurement
(see

“Stand- Alone DMM Mode”

below).

• To stop a scan in progress, press and hold the

Scan (Measure) key.

• To view the readings in memory, use the

View

key (the readings are not

erased when you read them).

Remote Interface Operation:

• To define the list of channels to be included in the scan list, use the

ROUTe:SCAN

command.

• To configure the measurement parameters on the desired channels,

use the

CONFigure

and

SENSe

commands.

• To initiate a scan and store all readings in memory, use the

INITiate

or

READ?

command. Each time you initiate a new scan, the instrument

will clear the previous set of readings from memory.

• To stop a scan in progress, use the

ABORt

command.

• To view the readings in memory, use the

FETCh?

command

(the readings are not erased when you read them).

N O T E

You can use the READ? command in one of three forms depending on
which measurement mode you wish to use.

•

If you omit the optional <ch_list> parameter and a scan list is not
currently defined, the READ? command applies to the internal DMM.

•

If you omit the optional <ch_list> parameter and a scan list is currently
defined, the READ? command performs a scan of the channels in the
scan list.

•

If you specify a <ch_list>, regardless of whether a scan list is currently
defined, the READ? command performs a “temporary” scan of the
specified channels (independent of the present scan list).

N O T E

You can use the MEASure? command in one of two forms depending on
which measurement mode you wish to use.

•

If you omit the optional <ch_list> parameter, the MEASure? command
applies to the internal DMM.

•

If you specify a <ch_list>, the MEASure? command performs a
“temporary” scan of the specified channels (independent of the
present scan list).

Advertising