66

34980A User’s Guide

2

Features and Functions

Self-Test

A power- on self- test occurs automatically when you turn on the
instrument. This limited test assures you that the instrument and all
installed plug- in modules are operational. This self- test does not perform
the extensive self test described below.

A complete self- test actually performs a series of internal tests and takes
approximately 20 seconds to execute. If all tests pass, you can have high
confidence that the instrument and all installed plug- in modules are
operational. This feature is available from the remote interface only.

• If you have a 34951A Isolated DAC Module installed, the complete

self- test will require an additional 15 seconds to complete per DAC
module (a memory test is performed).

• The complete self- test will abort if any signals are connected to ABus1

via the rear- panel Analog Bus connector (pins 4, 5, and 9; see

“Analog

Buses”

on page 15). Be sure to disconnect any signals from ABus1 prior

to running the self- test.

• If the power- on or complete self- test fails, and error is stored in the

error queue. See the Agilent 34980A Service Guide for more
information on returning the instrument to Agilent for service.

• Following the complete self- test, the instrument issues a Factory Reset

(

*RST

command).

Remote Interface Operation:

The following command returns “+0” if the

self- test is successful or “+1” if it fails.

*TST?

Front-Panel Display Control

For security reasons or for a slight increase in measurement rates,
you may want to turn off the front- panel display. From the remote
interface, you can also display up to 18 characters on the upper line
of the front- panel display.

• You can disable the front- panel display only by sending a command

from the remote interface (i.e., you cannot disable the front panel while
in local operation).

• When disabled, the entire front- panel display goes dark and all display

annunciators except ERROR, HOT, and Safety Interlock are disabled.

• The front- panel display is automatically enabled when power is cycled,

after a Factory Reset (

*RST

command), or after an Instrument Preset

(

SYSTem:PRESet

command).

Advertising