Measurement functions – Agilent Technologies Switch/Measure User Manual

Page 28

Advertising
background image

16

34980A User’s Guide

2

Features and Functions

Measurement Functions

The following table shows which DMM measurement functions are
supported by each of the multiplexer modules.

Note that similar considerations must be taken into account on the
34931A, 34932A, and 34933A matrix modules. Since the matrix modules
cannot be incorporated into a scan list, you must use the Stand- Alone
DMM Mode for these modules.

Front Panel Operation:

DMM or Channel (Configure) > DMM MEASUREMENT

Use the knob (or numeric keypad) to select the desired channel. Then
select the desired measurement function for this channel. You are
automatically guided to the next level of the menu where you can
configure other measurement parameters (range, integration time, etc.).

Remote Interface Operation:

You can select the measurement function using

the

CONFigure

and

MEASure?

commands. For example, the following

command configures the specified channel for dc voltage measurements.

CONF:VOLT:DC 10,DEF,(@3001)

Function

34921A

40-Ch Arm

MUX

34922A

70-Ch Arm

MUX

34923A

40-Ch Reed

MUX

(2-Wire)

34923A

80-Ch Reed

MUX

(1-Wire)

34924A

70-Ch Reed

MUX

34925A

40-Ch FET

MUX

(2-Wire)

34925A

80-Ch FET

MUX

(1-Wire)

Voltage, AC/DC

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Current, AC/DC

Yes

1

No

No

No

No

No

No

Frequency/Period

Yes

Yes

Yes

Yes

Yes

Yes

Yes

Ohms 2-Wire

Yes

Yes

Yes

5

Yes

5

Yes

5

Yes

6

Yes

6

Ohms 4-Wire

Yes

Yes

Yes

5

No

Yes

5

Yes

6

No

Thermocouple Yes

2

Yes

3

Yes

3,4

Yes

3,4

Yes

3,4

Yes

3

Yes

3

RTD 2-Wire

Yes

Yes

Yes

5

Yes

5

Yes

5

No

No

RTD 4-Wire

Yes

Yes

Yes

5

No

Yes

5

Yes

6

No

Thermistor Yes

Yes

Yes

5

Yes

5

Yes

5

No

No

1

Direct current measurements are allowed on channels 41 through 44 only (for all other channels, external shunts are required).

2

Optional 34921T Terminal Block is required for thermocouple measurements with built-in internal reference junction.

3

A fixed or external reference junction temperature is required for thermocouple measurement with this module.

4

Impact of higher offset voltage specification (< 50

µ

V) must be taken into consideration.

5

1 k

or higher range used unless 100

series resistors are bypassed on module.

6

10 k

or higher range used for loads over approximately 300

due to series resistance of FET channels.