Temperature Measurement Configuration

This section contains information to help you configure the instrument
for making temperature measurements. The table below shows the
thermocouple, RTD, and thermistor types for which the instrument
supports direct measurements.

Measurement Units

The instrument can report temperature measurements in °C (Celsius),

°F (Fahrenheit), or K (Kelvins). You can mix temperature units on
different channels within the instrument and on the same module.

The

CONFigure

and

MEASure?

commands automatically select °C.

Setting the Mx+B measurement label to °C, °F, or K has no affect on

the temperature measurement units currently selected.

The instrument selects Celsius when the probe type is changed and

after a Factory Reset (

*RST

command). An Instrument Preset

(

SYSTem:PRESet

command) or Card Reset (

SYSTem:CPON

command) does

not change the units setting.

Front Panel Operation:

DMM or Channel (Configure) > TEMPERATURE > UNITS

Remote Interface Operation:

UNIT:TEMP {C|F|K}[,(@<ch_list>)]

Temperature transducers supported

Thermocouple Types

*

* Using ITS-90 software conversions.

RTD Types

Thermistor Types

B, E, J, K, N, R, S, T

R

0

= 49

Ω to 2.1 kΩ

α = 0.00385 (DIN/IEC 751)

*

α = 0.00391

† Using IPTS-68 software conversions.

2.2 k

Ω, 5 kΩ, 10 kΩ

(YSI 44000 Series)