Temperature measurement configuration, Measurement units – Agilent Technologies Switch/Measure User Manual
Page 42

30
34980A User’s Guide
2
Features and Functions
Temperature Measurement Configuration
This section contains information to help you configure the instrument
for making temperature measurements. The table below shows the
thermocouple, RTD, and thermistor types for which the instrument
supports direct measurements.
Measurement Units
• The instrument can report temperature measurements in °C (Celsius),
°F (Fahrenheit), or K (Kelvins). You can mix temperature units on
different channels within the instrument and on the same module.
• The
CONFigure
and
MEASure?
commands automatically select °C.
• Setting the Mx+B measurement label to °C, °F, or K has no affect on
the temperature measurement units currently selected.
• The instrument selects Celsius when the probe type is changed and
after a Factory Reset (
*RST
command). An Instrument Preset
(
SYSTem:PRESet
command) or Card Reset (
SYSTem:CPON
command) does
not change the units setting.
Front Panel Operation:
DMM or Channel (Configure) > TEMPERATURE > UNITS
Remote Interface Operation:
UNIT:TEMP {C|F|K}[,(@<ch_list>)]
Temperature transducers supported
Thermocouple Types
*
* Using ITS-90 software conversions.
RTD Types
Thermistor Types
B, E, J, K, N, R, S, T
R
0
= 49
Ω to 2.1 kΩ
α = 0.00385 (DIN/IEC 751)
α = 0.00391
†
† Using IPTS-68 software conversions.
2.2 k
Ω, 5 kΩ, 10 kΩ
(YSI 44000 Series)