2

34980A User’s Guide

51

Sample Count

The sample count sets the number of auto- triggered samples the internal
DMM will take per channel per trigger. The sample count applies to both
scanning and stand- alone DMM measurements (with no scan list). The
front- panel sample annunciator (“ * ”) turns on during each measurement.

Sample count for Stand-Alone DMM Mode

Sample count for Scanning Mode

t

Sample Count

(1 to 500,000 samples)

Trigger

Trigger

t

Sweep 1

Trigger

Sweep 2

Sweep

n

Trigger

. . .

Sweep Count

Ch 1

Ch 3

Ch 4

Ch 2

Ch 5

Ch 6

Sample Count
(1 to 500,000 samples)