Section 2. installation and hardware set-up, 1 setting the sdm address – Campbell Scientific SDM-SIO4 4-Channel Serial I/O Interface User Manual

Page 13

Advertising
background image

2-1

Section 2. Installation and Hardware
Set-Up

The

SDM

-

SIO4

is

designed

to

be

mounted

on

an

enclosure

chassis

plate

using

the

two

mounting

holes

in

the

tabs

on

the

side

of

the

case

.

Before installation, it is necessary to set up the address of the SDM-SIO4 and also the
mode of operation of each serial port. These settings are determined by jumpers inside the
case. To access these, remove the four M3 screws and lift the lid off. With the connectors
facing you, you will see two blocks of jumpers on the right hand side of the circuit board.

Turn off the 12V supply and take static prevention
precautions before removing the lid.

2.1 Setting the SDM Address

The 4 x 2-way block of jumpers close to the ‘D’ connectors selects the SDM
address of the SDM-SIO4. This address ranges from 0 to 15. The four selector
blocks are numbered on the PCB silk screen as 1, 2, 4, and 8; ‘8’ is closest to
the ‘D’ connectors. This is in binary format as shown in Table 2-1.

TABLE 2-1. Address Settings

Selector Block Settings

SDM Address

8

4

2

1

0

0

0

0

0

0

0

0

1

1

0

0

1

0

2

0

0

1

1

3

0

1

0

0

4

0

1

0

1

5

0

1

1

0

6

0

1

1

1

7

1

0

0

0

8

1

0

0

1

9

1

0

1

0

10

1

0

1

1

11

1

1

0

0

12

1

1

0

1

13

1

1

1

0

14

1

1

1

1

15

NOTE: A ‘1’ means the selector block shorts both pins. ‘0’ means the selector
is only connected to one pin (for storage).

CAUTION

Advertising