Section 5. Programming the Datalogger

5-23

A5 – this filter sets up a 250ms filter time-out; this time-out should be long
enough for the sensor to transmit all of its data.

x – this filter marks the start of the data set. In this case the data set is two
floating point numbers.

f – this filter searches for the first ASCII floating point number to convert.

f – this filter searches for the second ASCII floating point number to convert.

*Table 1 Program

01: 2

Execution Interval (seconds)

;do filter set-up only if flag one is clear.

1: If Flag/Port (P91)

1:

21

Do if Flag 1 is Low

2:

30

Then Do

;set up string filter.

2: SDM-SIO4 (P113)

1:

1

Reps

2:

0

Address

3:

1

Send/Receive Port 1

4:

2054

Command

5:

9200

1st Parameters

6:

0

2nd Parameters

7:

0

Values per Rep

8:

0

Loc [ _________ ]

9:

1.0

Mult

10:

0.0

Offset

;delay required by filter set-up command.

3: Excitation with Delay (P22)

1:

1

Ex Channel

2:

0

Delay W/Ex (units = 0.01 sec)

3:

1

Delay After Ex (units = 0.01 sec)

4:

0

mV Excitation

;set flag one high so the filter set-up is only done once.

4: Do (P86)

1:

11

Set Flag 1 High

;end of IF.

5: End (P95)

;poll SDM-SIO4 to see if data is available.