Agilent Technologies Agilent 4396B User Manual
Page 44

Selecting
the
Input
The
analyzer
has
four
inputs;
S,
R,
A,
and
B
.
In
most
spectrum
measurements
,
the
S
input
is
used.
The
R,
A,
and
B
inputs
can
also
be
used
for
a
spectrum
measurement,
but
the
dynamic
range
of
these
inputs
is
20
dB
worse
than
the
S
input
and
the
attenuator
is
not
variable
.
Therefore
,
to
get
the
most
accurate
results
,
you
should
use
the
S
input
for
spectrum
measurements
.
In
the
spectrum
analyzer
mode
,
the
S
input
is
selected
by
default.
In
the
following
steps
,
you
verify
the
S
input
is
selected.
In
the
MEASUREMENT
block,
press
4
Meas
5 .
V
erify
the
F
FFF
S
in
F
FFFFFFFFFFFFFFFFFFFFFFFFFFFF
SPECTRUM:
S
is
underlined.
(This
shows
that
the
S
input
is
selected
for
a
spectrum
analyzer
measurement.)
Setting
the
Frequency
Range
The
CAL
OUT
signal
(20
MHz
at
020
dBm)
is
connected
as
test
signal
source
.
T
o
see
this
signal
on
display
,
you
must
set
the
appropriate
frequency
range
(in
this
case
,
0
to
80
MHz):
In
the
SWEEP
block,
press
4
Start
5 .
Press
4
0
5 .
2-6
Spectrum
Analyzer
T
our