Calculate:marker subsection – Agilent Technologies N9340A User Manual
Page 107

SCPI Command Reference
6
N9340A User’s Guide
105
CALCulate:MARKer Subsection
Markers All Off on All Traces
:CALCulate:MARKer:AOFF
This command turns off all markers on all the
traces.
Markers All On
:CALCulate:MARKer:ALL
This command turns on all the markers and places
them on six highest peaks.
Frequency Counter Marker
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt
[:STATe] OFF|ON|0|1
:CALCulate:MARKer[1]|2|3|4|5|6:FCOunt
[:STATe]?
This command turns on or off the marker
frequency counter. To query the frequency counter,
use
:CALCulate:MARKer[1]|2|3|4|5|6:FRECount
:X? If the specified marker number is not on, it
will be turned on and becomes the active marker
(normal mode). Once the marker count is on, it is
on for any active marker, not just for the one used
Key Access:
[Marker] > {More} > {All Off}
*RST State:
Off
Key Access:
[Marker] > {Peak search} > {Auto search}