Single sweep – Agilent Technologies N9340A User Manual
Page 78

5
Key Reference
76
N9340A User’s Guide
• Fast
Activates the fast sweep mode.
Fast sweep mode provides a fast measurement
function which decreases the sweep time, but this
mode brings a decrease of amplitude accurate.
Key Access:
[BW/SWP] > {Sweep} > {Sweep Mode}
Single Sweep
When analyzer is in continuous sweep mode and
not in a measurement
([MEAS] > Measure Off),
Pressing
[BW/SWP] > {Single Sweep} convert the
continuous sweep to single sweep and executes a
sweep after the trigger condition is met. If the
analyzer is already in single sweep, pressing
[BW/SWP] > {Single Sweep} executes a new sweep
after the trigger condition is met.
If Average is on
([TRACE] > {More (1 of 2)} > {Average}
> On), pressing [BW/SWP] > {Single Sweep} halts the
average on the current average number, and
continue the average from the current count
number as soon as the single sweep function
triggers.
Key Access:
[BW/SWP] > {Sweep}