Measure subsystem, Obw subsection, Acpr subsection – Agilent Technologies N9340A User Manual
Page 121

SCPI Command Reference
6
N9340A User’s Guide
119
MEASure Subsystem
Provides programming information for the keys
associated with the measurements available when
you press the front- panel Measure key.
OBW Subsection
Setting Percentage (%) method of OBW
:MEASure:OBW:PERCent <value>
:MEASure:OBW:PERCent?
edit the percentage of signal power used when
determining the occupied bandwidth. Press
{%} to
set the percentage ranging from 10.00% to 99.99%.
Setting dBc method of OBW
:MEASure:OBW:XDB <value>
:MEASure:OBW:XDB?
specify the power level used to determine the
emission bandwidth as the number of dB down
from the highest signal point, within the occupied
bandwidth span.
ACPR Subsection
Center Freq
:MEASure:ACPR:CENTer <value>
:MEASure:ACPR:CENTer?
Sets the center frequency of the main channel
power.