Center frequency step size automatic, Center frequency step size, Frequency span – Agilent Technologies N9340A User Manual
Page 132

6
SCPI Command Reference
130
N9340A User’s Guide
Center Frequency Step Size Automatic
[:SENSe]:FREQuency:CENTer:STEP:AUTO
OFF|ON|0|1
[:SENSe]:FREQuency:CENTer:STEP:AUTO?
This command specifies whether the step size is
set automatically based on the span.
Center Frequency Step Size
[:SENSe]:FREQuency:CENTer:STEP
[INCRement] <freq>
[:SENSe]:FREQuency:CENTer:STEP
[INCRement]?
This command specifies the center frequency step
size.
Frequency Span
[:SENSe]:FREQuency:SPAN <freq>
[:SENSe]:FREQuency:SPAN?
This sets the frequency span.
Setting the span to 0 Hz puts the analyzer into
zero span.
*RST State:
On
Key Access:
[FREQ] > {CF Step}
*RST Value:
300 MHz
Available unit:
Hz, kHz, MHz, GHz
Key Access:
[FREQ] > {CF Step}
*RST Value:
3.0 GHz
Available unit:
Hz, kHz, MHz, GHz
Key Access:
[SPAN] > {Span}