Connections in the ip67 model – NOVUS LogBox-DA User Manual

Page 4

Advertising
background image

LogBox-DA

NOVUS AUTOMATION

4/7

Note: Make sure that the compress fitting is perfectly tightening the

cables, thus assuring proper IP65 protection (totally dust-tight and

protected against water jets).

Fig. 8 – IP65 connections internal view

CONNECTIONS IN THE IP67 MODEL
In theIP67 version, a M8 connector is provided for signal input, as in

Fig. 9. When required, a second M8 connector can be installed to

output the signal for commanding the external power supply. The

cables are supplied with the logger.
Note: The case cover should not be opened unless battery

replacement is required. If this is the case, the cover must be

properly tightened back in its place such as to assure the IP67 / 68

protection.

Fig. 9 – IP67 external connectors

Wire connections

Channels cable (right)

Connection

Blue

Channel 1 – (3 IN1) – positive (+)

Black

Channel 1 – (4 GND) – negative (-)

White

Channel 2 – (1 IN2) – positive (+)

Brown

Channel 2 – (2 GND) – negative (-)

Switch cable (left)

Connection

Blue

Transducer (7 +VTRANS)

Black

External Battery (5 -BAT)

White

External Battery (6 +BAT)

Brown

Not connected


Fig. 10 – Voltage pulse and dry contact channel 1 input

Fig. 11 – NPN and PNP open collector input for channel 1

Fig. 12 – Current and voltage inputs for channel 2