5. Operation

Before the pump can be operated, the pumping data must be setup. At minimum, the syringe
inside diameter and a non-zero pumping rate needs to be set. The operation of the pump can
then be started from the TTL I/O connector, or from RS-232 computer control.

All data that is set will be stored in the pump’s non-volatile memory, meaning that the new value will not be
lost the next time that power is applied to the pump. The only exception is when the pumping rate is changed
while the Pumping Program is operating. In this case the new pumping rate will not be stored in non-volatile
memory.

In general, numeric data are floating point data consisting of up to 4 digits plus 1 decimal point, and is referred
to in the command descriptions as <float> data. For example 123.4, 12.3 and 12.34 are acceptable formats for
floating point data.

5.1 Syringe Inside Diameter

Set with command: DIA <float>

The syringe inside diameter can only be set while the Pumping Program is stopped.

Valid syringe diameters are from 0.1 mm to 50.0 mm. If the diameter is out of this range, the response will be
‘?OOR’. Changing the syringe diameter will not zero any current settings. Section 10.7, “Syringe Diameters
and Rate Limits”, is a representative list, for reference, of syringe diameters for various syringe manufacturers
and syringe sizes.

5.1.1 Default Volume Units

The units of the accumulated infusion and withdrawal volumes and the “Volume to be Dispensed” are set
according to the diameter setting. NOTE: A change in the volume units will affect all “Volume to be
Dispensed” settings in the Pumping Program. If the default volume units are changed (see next section), the
selected volume units will remain in affect until a reset function is performed.

From 0.1 to 14.0 mm

Syringes smaller than 10 mL:

Volume units are ‘

L’

From 14.01 to 50.0 mm

Syringes greater than or equal to 10 mL: Volume units are ‘mL’

5.1.2 Changing Volume Units

The volume units used for accumulated volumes and the “Volume to be Dispensed” settings can be changed to
either ‘mL’ or ‘

L’. A change in the volume units will affect all “Volume to be Dispensed” settings in the

Pumping Program.

Set the volume units command:

VOL {ML | UL}

The selected volume units will remain in affect and override the default volume units. Changing the diameter
will no longer change the volume units. Performming a system reset (*RESET) will cancel the override and
allow the volume units to change to the default volume units when setting the syringe diameter.

5.2 Start/Stop Triggers

Start Command: RUN

Stop Command: STP

The Pumping Program can be started or stopped from different sources. The “RUN” and “STP” commands, or
the TTL I/O Operational Trigger input. Each can control the Pumping Program’s operation.

5.3 Operating the Pump

When a start trigger is given, the Pumping Program begins to operate, starting with Phase 1. If the current
Program Phase specifies a pumping rate, the pump will begin pumping. The pumping direction will also