Pump reset, Quick start instructions, Operating the pump – Next Advance SP2200 User Manual

Page 2

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe Pump

Publication #1200-02

i

04/29/10

Quick Start Instructions

Assumes that the pump was not previous programmed with a multiple Phase Pumping Program.

Plug in the pump.

Attach the computer cable to the "Computer" connection on the RJ-11 connector

Start the terminal emulator program on the computer (downloadable from the web site).

Baud Rate: 19,200
Data frame: 8N1

(Default settings for the downloadable terminal emulators)

Note: Each command sent to the pump must have a carriage return (0x0D) at the end, but space characters are
optional and will be ignored. (Sent automatically by terminal emulator).

Note: Wait for the response from each command before sending the next command

Setup Pumping Parameters

Set the Syringe Inside Diameter:

Where "nn.nn" is the syringe inside diameter in mm: Send the command: DIA nn.nn

Set the Pumping Rate.

Where "nn.nn" is the pumping rate and "xx" is the rate's units: Send the command: RAT nn.nn xx

Rate units "xx" value
mL/hr

MH

µL/hr

UH

mL/mn

MM

µL/mn

UM

If the pumping rate is out of range the response message to the computer will include "?OOR"

Set the Volume to be Dispensed or Continuous Pumping

Where "nn.nn" is the volume to be dispensed, send the command: VOL nn.nn

If the volume to be dispensed is 0, then the pump will pump continuously.

Note: The default units for the volume to be dispensed will be mL or µL, depending on the syringe size.
At or below 14.00 mm syringe diameter, volume units are µL.
Above 14.00 mm syringe diameter, volume units are mL. The default volume units can be overriden.

Set the Pumping Direction

For infusion, send the command:

DIR INF

For Withdrawal, send the command:

DIR WDR

Load the Syringe

Turn and lift center knob to release the pusher block, then slide to the required position.

Insert the syringe plungers in the pusher block slots.

Insert the syringe barrel flanges in the flange bracket, with the syringe barrel holder on the syringes. Tighten the
flange brackets onto the syringe flange. Tighten the pusher block screws. Turn knob to lock pusher block.

Operating the Pump

Start the Pump:

Send the command: RUN

Stop the Pump:

Send the command: STP

Query the Pump Status

Send just a carriage return to the pump. The pump will respond with a status prompt

When Pumping

The pumping rate can be changed.

With continuous pumping, the pumping direction can be changed.

PUMP RESET:

Send the command: *RESET