8.5.1 Program Function Commands

The following commands are relevant to the currently select Program Phase. Note: During a Pumping
Program’s operation, the currently selected Phase can change automatically.

PHASE NUMBER
PHN [ <phase data> ]

Set/query currently selected Program Phase:

Set:
Currently selected Phase is set to <phase data>. Previous Phase is stored in non-volatile memory and
the requested Phase is recalled from the non-volatile memory. Set is only valid if the Pumping
Program is not operating.

Query response:

<phase data>

Currently selected Phase

PUMPING PROGRAM FUNCTION
FUN [ <phase function> ]

Set/query the Pumping Program Phase’s function.

This command is relevant to the currently selected Phase. Set is only valid if the Pumping Program is
not operating

For a more detailed description of Program commands, see sec. 7.3, Program Function Descriptions.

<phase function> =>

Rate Data Functions

When a Phase’s function is set to a “Rate Data Function”, use the ‘RAT’, ‘VOL’, and ‘DIR’
commands to setup the pumping parameters for the current phase.

RAT

Pumping rate.

INC

Increment rate.

DEC

Decrement rate.

Non-Rate Control Functions

STP

Stop pump.

PAS <number data>

Pauses pumping for ‘nn’ seconds, or until trigger if 0.

PAS <n.n>

Pauses pumping for 'n.n' seconds.

LOP <count data>

Loop to previous loop start ‘nn’ times.

LPS

Loop starting Phase.

LPE

Loop end Phase.

JMP <phase data>

Jump to Program Phase.

Conditional Functions

IF

<phase data>

If Program input TTL pin low, jump to Phase.

EVN <phase data>

Set event trigger trap.

EVS <phase data>

Set event square wave trigger trap.

EPL <expansion pin>

Pause and wait for expansion pin event edge or level.

EPE <expansion pin>

Pause and wait for expansion pin event edge.

EVE <expansion pin>

Re-direct event trigger trap to expansion pin.

EVR

Event trigger trap reset.

TTL Logic and Other Functions

OUT <TTL level>

Set programmable output pin.

OE0 <expansion pin>

Set expansion pin to 0 (low level).

OE1 <expansion pin>

Set expansion pin to 1 (high level).

BEP

Sound short beep.

PUMPING RATE