Next Advance SP2200 User Manual

Page 25

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

20

04/29/10

Phase

Function

Rate

Volume

Direction

1

RAT

200 ml/hr

0.1 ml

Infuse

Phase

Function

2

LPS

Phase

Function

Rate

Volume

Direction

3

INC

1.0

0.1 ml

Infuse

Phase

Function

4

LOP 50

Phase

Function

5

LPS

Phase

Function

Rate

Volume

Direction

6

DEC

1.0

0.1 ml

Infuse

Phase

Function

7

L0P 99

Phase

Function

Rate

Volume

Direction

8

DEC

1.0

0.1 ml

Infuse

Phase

Function

9

LPS

Phase

Function

Rate

Volume

Direction

10

INC

1.0

0.1 ml

Infuse

Phase

Function

11

LOP 50

Phase

Function

12

JMP 02

7.4.4 Example 4: Complex Dispenses with External Synchronization

0

0

750

Syringe

Refill

A more complex dispensing example, this Program contains different pumping requirements, including
dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during
the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for
a start trigger before starting the next dispense.

The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
dispenses, the syringe is refilled by the amount infused, 17.25 ml. Then the buzzer beeps, to alert the operator
to the start of the first set of dispenses. The process is then repeated.

Advertising