2 rs-232 pump network connector wiring – Next Advance SP2200 User Manual

Page 47

Advertising
background image

NEXT>>>ADVANCE

www.nextadvance.com

SP2200Programmable Syringe

Pump

Publication #1200-02

42

04/29/10

*ADR [ DUAL | RECP ]

Set Reciprocating or Dual pumping mode

*RESET

Clears program memory and resets communication parameters
(system command, valid regardless of current address)

------------------------------------------------------------------------------------------------------------------------------
Network Command Burst => <n> <command> * [Network Command Burst]

Communicate with multiple pumps on a pump network simultaneously

10.2 RS-232 Pump Network Connector Wiring

RS-232

To Network To Computer

PC Com Port Connectors

25 Pin

9 Pin

3 - Receive

2 - Receive

Connect to pump TXD

2 - Transmit

3 - Transmit

Connect to pump RCV

9 - Ground

5 - Ground

Connect to pump GND

10.3 Accessories

10.3.1 ANA-BOX

Part #: ADPT-ANABOX, Analog voltage control interface.

Allows the pump to be controlled by a variable voltage source. Allows the Creation of a closed loop system
allowing the NE-1000 to be controlled by an external sensor, such as a pressure sensor, or other variable
voltage source. Start or stop the pump at a specific voltage level. Set the pumping rate to be proportional to
the voltage input.

10.3.2 Syringe Heater

Part#: SYRINGE-HEATER, see www.SyringeHeater.com for details.

Flexible heating pad that wraps around the syringe. Digital PID controller will heat a syringe to a set
temperature up to 100 C.

10.3.3 RS-232 Network Cables

RS-232 Network Primary Cable
7 foot cable, part #:

CBL-PC-PUMP-7

25 foot cable, part #:

CBL-PC-PUMP-25

RCV from network

TXD to network

TXD to computer

RCV from computer