FMI PDS100 User Manual

Page 2

Advertising
background image

Page 2 of 18

Smooth Flow Operation

The PDS Smooth Flow product differs from the “Standard” PDS-100 in

that the dual pumps have each been precisely calibrated to provide

equal dispenses per revolution. The distinct difference found in the PDS

Smooth Flow is in the method by which the pumps are driven.

Proprietary programming provides a unique method of control in such a

way as to provide a virtually pulseless output when both pumps are

connected to a “Y” or “Tee” fitting.

Most of the controls and screens are the same for the Smooth Flow as

shown in the previous pages with the exception that the pumps are not

individually controllable. Instead the pumps are controlled as though

they are one.

Menus, for the most part, are the same except while in dispense or

continuous mode the word “Smooth” will appear at the top of the main

screen. In addition you will see the word “Smooth Drive” when power is

first applied to the unit.










Fuse Location

Power Switch

FEATURES:

“Set and Forget” approach

1-9999 dispenses (adjustable)

Table top/wall mount

Universal power supply with standard IEC line

cord

CSA/UL, CE, RoHS compliant

Learn mode/count to allow customer to command

the PDS-100 to “remember” desired dispense

cycles needed to fill to a desired volume.

Prime Mode

Purge/Reverse Mode

Table-top Mounting: The system is configured for table top

installation initially with rubber grommets slid into the mounting slots.

Before moving on to configuring the control module, make sure that

these rubber grommets are in place. There are no further installation

steps for table top use.

Wall-Mounting: For wall mounting, it is necessary to remove the

four rubber grommets from the control module before attempting to

mount. Each unit must be mounted in the correct orientation; with the

labels facing right side up, the control module will have the pumps at

the bottom. Wall mounting may require an appropriate mounting

board of at least 1/2”(12mm) thickness to straddle the studs of a

typical plasterboard wall. See Figure 3.

IMPORTANT When mounting PDS-100 in an enclosure it is

important to maintain a minimum clearance to provide adequate

space to reach the power switch. See figure 2.

Figure 1

PDS-100 Rear View

Figure 2

Minimum Power Switch Clearance

Figure 3

Required Mounting Pattern

Specifications

Supply Voltage: 100-240VAC +/- 10%, 50/60 Hz. Main Supply

Current: 0.6 to 0.25 Amps. Fuses: T250V-1A (time lag), 5x20mm, 2

required.

Physical: Dimensions: 5" H x 6 1/4" W x 12" (max) D; Weight: 7 lbs

Environment: For indoor use only. Humidity: 80% max for

temperatures up to 31°C, decreasing linearly to 50% relative humidity

at 40°C. Pollution degree, 2. Operating temperatures range from 5ºC

to 40ºC (41ºF to 104ºF).

4.0”

MINIMUM SIDE

CLEARANCE

REQUIRED

4.0”

MINIMUM REAR

CLEARANCE

REQUIRED

Switch Clearance

Rear View

IN-PDS100-13A

11/1/13 HW