INSTALLATION - SITE REQUIREMENTS

Dental Site Requirements

JVac3

JVac5

JVac7

Electrical
Wire gauge (AWG)

#10

#10

#10

Minimum circuit breaker current rating

30 A

30 A

30 A

Plumbing
Exhaust vent pipe to outside

1.5 inch NPT (must be metal)

Line from operatory

See applicable plumbing diagram

Inlet hose barb on vacuum separator tank

1.5 inches

1.5 inches

1.5 inches

Separator tank drain hose

1.5 inches

1.5 inches

1.5 inches

Note: if alternate site pipe sizes are to be used, consult an authorized distributor

Minimum 12" spacing

between system & ceiling

(recommended 18" for servicability)

1-1/2" NPT

system drain

NOTE: All dimensions are in inches

Minimum 6" spacing

between system & any walls

Inlet from operatory

for 1-1/2" I.D. tube

1-1/2" NPT

system exhaust

Ø 15.25

0.7

63.4

32.5

61.9

21.29

24.1

6 in min.

13.4

26.0

12 inches min.; 18 inches recommended