Installation - site requirements – JUN-AIR JVac Dental Vacuum Suction System User Manual
Page 10
10
CI-100A (Rev B)
JUN-AIR Dental Vacuum Suction System User Guide
© 2012, JUN-AIR
We reserve the right to make any alterations which may be due to any technical improvements
Printed in the USA
INSTALLATION - SITE REQUIREMENTS
Dental Site Requirements
JVac3
JVac5
JVac7
Electrical
Wire gauge (AWG)
#10
#10
#10
Minimum circuit breaker current rating
30 A
30 A
30 A
Plumbing
Exhaust vent pipe to outside
1.5 inch NPT (must be metal)
Line from operatory
See applicable plumbing diagram
Inlet hose barb on vacuum separator tank
1.5 inches
1.5 inches
1.5 inches
Separator tank drain hose
1.5 inches
1.5 inches
1.5 inches
Note: if alternate site pipe sizes are to be used, consult an authorized distributor
Minimum 12" spacing
between system & ceiling
(recommended 18" for servicability)
1-1/2" NPT
system drain
NOTE: All dimensions are in inches
Minimum 6" spacing
between system & any walls
Inlet from operatory
for 1-1/2" I.D. tube
1-1/2" NPT
system exhaust
Ø 15.25
0.7
63.4
32.5
61.9
21.29
24.1
6 in min.
13.4
26.0
12 inches min.; 18 inches recommended