JUN-AIR JVac Dental Vacuum Suction System User Manual
Page 19
19
© 2012, JUN-AIR
We reserve the right to make any alterations which may be due to any technical improvements
Printed in the USA
JUN-AIR Dental Vacuum Suction System User Guide
CI-100A (Rev B)
Specifications
JVac3
JVac5
JVac7
Electrical
Voltage
230
230
230*
Frequency (Hz)
60
60
60
Phase
Single
Single
Single
Operating current (amps)
6.3 A
12.5 A
12.5 A
Starting current (amps)
50 A
70 A
70 A
Horsepower
1.5
3.0
3.0
Insulation class
F
F
F
Pneumatic
Max. vacuum (in Hg)
8.5
10
12
Vacuum limiter preset (in Hg)
7.25
9
10
Open vacuum air flow (CFM)
60
80
100
Separation tank capacity
(working fluid)
8 gallons
8 gallons
8 gallons
Tank material
stainless steel
stainless steel
stainless steel
Ambient Specifications
Operating temperature
4.5 °C to 40 °C [40 °F to 104 °F]
Relative humidity
(non-condensing)
0 to 90 %
0 to 90 %
0 to 90 %
Environment
clean and dust free
clean and dust free
clean and dust free
Dimensions
Height (in)
64
64
64
Diameter (in)
26
26
26
Weight (lb)
190
225
225
* See page 15 ‘INSTALLATION - ELECTRICAL’.