PERSONAL SAFETY:
DANGER: Danger of fire or explosion
when using flammable substances. Do
not operate the vacuum system in an
area containing combustible gases or
anesthetic mixtures.
CAUTION: Do not place vacuum system
within 6 feet of patient.
CAUTION: Never leave children
unattended near vacuum system when
in use.
WARNING: Property damage and/or
personal injury may result if directions
are not followed or manufacturer’s
replacement parts/accessories are not
used.
WARNING: Connect only equipment
suitable for listed maximum vacuum
level of the blower. Equipment is not
suitable for use in vital (continuation of
life) operations, an alternative air source
should be used in these instances.
WARNING: A leaking vacuum relief valve
may indicate a need for adjustment
or repair. Any required adjustments
are only to be made by an authorized
distributor. Improper adjustment will
cause overheating and failure.
PLACEMENT:
• Indoor use only in dust free, climate controlled
room.
• DO NOT install in an enclosed area where
ambient temperature could exceed temperature
specifications of below 4.5 °C / 40 °F, or above
40 °C / 104 °F while the vacuum system is run-
ning.
• Maintain minimum 6 inches of clearance around
system. Maintain 12 inches of clearance on top
of system.
• Vacuum systems are equipped with 4 adjust-
able feet, to ensure unit stands firmly on the floor.
Level the system in two plains prior to starting
system.
• Assure proper external exhaust ventilation prior to
starting the system (see SPECIFICATIONS).
INSTALLATION SAFETY DATA
Indicates the ON and OFF position for
the equipment power switch (system
breaker)
When
ON, the indicator light will
illuminate and voltage
WILL be
supplied to system.
When
OFF, the indicator light will
NOT illuminate and voltage
WILL NOT
be supplied to the system.
I
o
= ON
= OFF
I
o
Equipment Alert: Vacuum system
must be installed per local plumbing
and electrical codes.