INSTALLATION - ABOVE GRADE PLUMBING

Installation Guidelines

• Use 1.5 inch to 2.0 inch ID plumbing to connect

operatory lines to separation tank.

• Plumbing lines should be sloped down to the

separation tank as shown in the schematic
(minimum slope of 1/32 inch per foot).

• A maximum of a 10-foot rise is recommended

for above-grade plumbing. If more than 10 feet
of rise is required, contact your distributor.

• Low spots in the line should be avoided to

prevent flow restriction or vacuum loss.

• It may be necessary to upgrade to the next

largest system for an operatory with above-grade
plumbing.

• All plumbing entering the main trunk line needs

to be dropped in on the top of the trunk line as
shown in the schematic.

the separation tank

Minimum 1.5" ID Pipe

Lines to Operatory

Slope pipe down towards

Ground Level

Max 10

Feet Height

Connection must
be higher than
tank inlet