JUN-AIR Dental Vacuum Suction System User Guide

CI-100A (Rev B)

INSTALLATION - BELOW GRADE PLUMBING

Installation Guidelines

• Where possible, use 1.5 inch to 2 inch

ID plumbing to connect operatory lines to
separation tank.

• Plumbing lines should be sloped down to the

separation tank as shown in the schematic
(minimum slope of 1/32 inch per foot).

• Low spots in the line should be avoided to

prevent flow restriction of vacuum loss.

the separation tank

Minimum 1.5" ID Pipe

Risers to

Operatory

Risers to Operatory

Slope pipe down towards

Ground Level

Connection must
be higher than
tank inlet