JUN-AIR Dental Vacuum Suction System User Guide
CI-100A (Rev B)
INSTALLATION - BELOW GRADE PLUMBING
Installation Guidelines
• Where possible, use 1.5 inch to 2 inch
ID plumbing to connect operatory lines to
separation tank.
• Plumbing lines should be sloped down to the
separation tank as shown in the schematic
(minimum slope of 1/32 inch per foot).
• Low spots in the line should be avoided to
prevent flow restriction of vacuum loss.
the separation tank
Minimum 1.5" ID Pipe
Risers to
Operatory
Risers to Operatory
Slope pipe down towards
Ground Level
Connection must
be higher than
tank inlet