Maintenance and cleaning

To avoid contamination, it is essential to check and change
the bath fluid from regularly.
If water is used for bath fluid, we recommend to add Wa-
ter bath protective media. The protective media stops the

growth of algae, bacteria and other microorganisms with
its bactericidal effectiveness. It protects the bath fluid and
provides a long period of use for the fluid.

To maintain the full cooling performance, maintain the
condenser dust filter from time to time.
• Switch off the device and disconnect mains cable.
• Open the front venting grid.
• Remove the bracket (see Fig. 7).

• Clean condenser dust filter with a vacuum cleaner or wash

the filter in the water and dry the it before assembly.

Don’t touch the heating exchang-
er surface with hard parts.

For cleaning disconnect the main
plug!

Use only cleaning agents which have been approved by
IKA

®

to clean the device.

Dirt

Cleaning agent

Dye

Isopropyl alcohol

Construction material Water containing tenside/isopropyl alcohol
Cosmetics

Water containing tenside/isopropyl alcohol

Foodstuff

Water containing tenside

Fuel

Water containing tenside

For materials which are not listed, please request information
from IKA

®

application support.

Wear protective gloves while cleaning the instruments.
Electrical instruments may not be placed in the cleansing
agent for the purpose of cleaning.
Do not allow moisture to get into the instrument when
cleaning.
Before using another than the recommended method for
cleaning or decontamination, the user must ascertain with
IKA

®

that this method does not damage the instrument.

Spare parts order:

When ordering spare parts, please give:
- machine type
- serial number, see type plate
- item and designation of the spare part,
see www.ika.com, spare parts diagram and spare parts list

- Software version.

Repair:

Please send in instrument for repair only after it

has been cleaned and is free from any materials

which may constitute a health hazard.
For repair, please request the “Decontamination Cer-

tificate” from IKA

®

, or download printout of it from the

IKA

®

website www.ika.com.

If you require servicing, return the instrument in its original
packaging. Storage packaging is not sufficient. Please also
use suitable transport packaging.