IKA RC 2 basic User Manual

Page 8

Advertising
background image

8

• Fluids (Standard information for IKA

®

fluid):

No.

IKA

®

Designation

Operating temperature

range for open bath

application

(°C)

Operating temperature

range for closed bath

applications

(°C)

Safety temperature

(°C)

Flash point

(°C)

0

CF.EG28.N10.80.8

-10 ... 80

-10 … 80

80

115

1

CF.EG39.N20.80.16

-20 … 80

-20 … 80

80

115

2

CF.EG44.N25.80.19

-25 … 80

-25 … 80

80

115

3

CF.EG48.N30.80.22

-30 … 80

-30 … 80

80

115

4

UF.Si.N30.150.10LV

-30 … 130

-30 … 150

145

>170

5

HF.Si.20.200.50

20 … 200

20 … 200

255

>280

6

HF.Si.20.250.50A

20 … 200

20 … 250

255

>280

7

Water

5 … 95

5 … 95

-

-

8

Customized

The listed fluid temperature range may deviate from the device operating temperature ranges and settings.

Note: for open bath application!

Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.

High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0.1 g soda (Na

2

CO

3

,

sodium carbonate) per liter of water.

Nomenclaturefor IKA

®

fluids:

CF.EG28.N10.80.8 --

(1) Classification:

HF: Heating Fluid

CF: Cooling Fluid

UF: Universal Fluid

(2) Chemical composition:

Si: Silicone oil

EG: Ethylene Glycol

(3) Temperature range: (Minimum temperature. Maximum temperature)

N: Negative Temperature

(4) Viscosiy:

Viscosiy at 25 °C for Heating Fluid (HF)

Viscosity at -20 °C for Cooling Fluid (CF)

Viscosity at 25 °C for Universal Fluid (UF)
Dynamic viscosity [mPa.s] is a product of kinematic viscosity [mm²/s] and density [kg/m³] of the fluid divided by 1000.

(5) Additional information:

A: Oil Additives

LV: Low Viscosity

(5) Addition information

(4) Viscosity (8 mm²/s)

(3) Temperature range (-1080 °C)

(2) Chemical composition (Ethylene Glycol 28%)

(1) Classification (Cooling fluid)