5

• A laboratory circulator refrigerates and circulates fluid

according to specified parameters. This involves hazards
due to low temperatures and general hazards due to the
device of electrical energy. The user is largely protect-
ed by the device of relevant standards. Further hazard
sources may arise due to the type of tempering fluid,
e.g. by exceeding or undercutting certain temperature
thresholds or by the breakage of the container and reac-
tion with the refrigeration carrier fluid. It is not possible
to consider all eventualities. They remain largely subject
to the judgment and responsibility of the operator. For
this reason, it may become necessary for user to take
other precautionary safety measures.

• Insufficient ventilation may result in the formation of ex-

plosive mixtures. Only use the device in well ventilated
areas.

• Use suitable hoses for connection.
• Secure hoses and tubes against slippage and avoid kinks.
• Check hoses, tubes and bath at regular intervals for pos-

sible material fatigue (cracks/leaks).

• When device is used for external circulation, extra pre-

caution must be taken for cold fluid leakage due to
damaged hose.

Do not start up the device if:

- It is damaged or leaking
- Cable (not only supply cable) is damaged.

• After a power failure during operation, the device may

start automatically (depending on operating mode).

• Transport the device with care.
• Do not transport or empty the bath while it is still cold.

Check the temperature of the bath fluid before draining.

• Always empty the bath before moving the unit.

Always empty the bath if the unit
is unused for long time.

Fluids:

Only use the fluids, which fulfill the
requirements for safety, health and
device compatibility. Be aware of

the chemical hazards that may be associated with the
bath fluid used. Observe all safety warning for the fluids.

• Depending on the bath fluid used and the type of opera-

tion or toxic can arise. Ensure suitable extraction.

• Do not use any fluid which may cause dangerous reac-

tions during processing.

• Only use recommended bath fluid. Only use non-acid

and non corroding fluid.

Never operate the device without
sufficient fluid! Check the fluid lev-
el detection at a regular basis (see

the section "Filling and draining").

• Continuous monitoring of the filling level of the bath fluid

is required.

• For optimum temperature stability, the fluids viscosity

should be 50 mm

2

/s or less at its lowest operating tem-

perature. This permits good fluid circulation and mini-
mizes heating from the pump.

• Untreated tap water is not recommended. It is recom-

mended to use distilled water or high purity water (ion
exchangers) and add 0.1 g soda (sodium carbonate
Na

2

CO

3

) /liter, to reduce corrosive properties.

Don't use following fluids:

- Untreated tap water

- Acids or bases
- Solutions with halides: chlorides, fluorides, bromides,

iodides or sulfur

- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.

DANGER

NOTICE

CAUTION

NOTICE

NOTICE

Correct use

• Use:

Use RC (Refrigerated Circulators) for cooling and circulat-

ing fluids.

Intended Use: Tabletop device

• Range of use (indoor use only):

- Laboratories

-Schools

- Pharmacies

- Universities

This instrument is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply

network that also supplies residential areas.

The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not

supplied or recommended by IKA

®

.

- If the device is operated improperly or in contrary to the

IKA

®

specifications.

- If the device or the printed circuit board are modified by

third parties.

Advertising