Technical data – IKA RC 2 basic User Manual

Page 17

Advertising
background image

17

Technical data

Nominal voltage

VAC

230 ± 10% / 115 ± 10%

Frequency

Hz

50 / 60

Max. input power

W

300

Working temperature range

°C

-20 ... RT

Operating temperature range (with external heating) °C

-20 ... + 40

Temperature stability – internal temperature
control (according to DIN 12876)

K

± 0.1

Temperature control

PID

Temperature measurement, absolute accuracy

K

± 0.5

Temperature setting

Key on the front of the device

Temperature setting resolution

K

0.1

Temperature display

LED

Temperature display resolution

K

0.1

Classification according to DIN 12876-1

Class III (FL) suitable for flammable and non-flammable fluids

Cooling capacity according to DIN 12876

+ 20 °C

+ 10 °C

0 °C

- 10 °C

- 20 °C

W

400
370
320
240
130

Pump speed (adjustable)

rpm

1000 ... 3200

Max. pump pressure/suction

bar

0.3/0.2

Max. flow rate (at 0 bar)

l/min

15

Bath volume

I

1.4 – 4.0

Fluid maximum viscosity

mm

2

/s 50

Sub-level protection

Yes

Interface

USB, RS 232

Permitted on-time

%

100

Protection class according to EN 60 529

IP 21

Protection class

I

Excess voltage category

II

Contamination level

2

Permitted ambient temperature

°C

+ 5 ... + 32

Permitted ambient humidity

%

80

Dimension (W x D x H)

mm

220 x 525 x 475

Weight

kg

28

Operation at a terrestrial altitude

m

max. 2000

Subject to technical changes!