IKA RV 8 FLEX User Manual

Page 26

Advertising
background image

26

Glassware
NOTE:
Please read the operating instructions of the glassware for
the safe handling of laboratory glassware !
• Open the locking device (O) on the drive head by turning it 60°

counterclockwise. (Fig.14a)

• Feed the vapour tube in until it stops.
• Then, lock the locking device (O) by turning it clockwise by 60°.

(Fig. 14b)

• The vapour tube is not allowed to be pulled out !
• Check the correct axial locking device on the vapour tube.

O

R

Fig.14b

O

Fig.14a

See indication:

“CLOSE„

Mount the evaporation flask:

- Place the evaporation flask (R) on the Vapour tube. (Fig.15)

- Turn the plastic nut (P) with the clip (S) counterclockwise until

the clip can be put on the flange. (Fig. 16)

- Now turn the plastic nut clockwise until the clamp is hard up

against the flange. When doing so, hold the locking device (O)

on the steam pipe.

Remove the evaporation flask:

- Hold the locking device (O) and loosen the plastic nut by turning

it counterclockwise. This releases a tightly clamped evaporating
flask.

- Hold the evaporating flask at the flange and open the jaws of

the clamp.

- Remove the evaporating flask.
- Check that the locking of vapour tube is still closed!

S

O

P

Fig.16

Turn clockwise () for tight-fitting
Turn counterclockwise (
) for loosen tight-fitting

Fit & unfit the evaporation flask / push off mechanism

O

R

P

Fig.15