• RV10.5 Vertical-condenser with manifold and cut-off valve

for reflux distillation

Also available with coating (RV10.50)

• RV10.6 Vertical-intensive condenser with manifold and cut-

off valve for reflux distillation

This vertical-intensive action condenser features a double jacket
design for particularly efficient reflux distillation.
Also available with coating (RV10.60)

Hose system

• Connect the water hoses (H

2

O) to the condenser according to

the counter-flow principle. (Fig. 20)

• Install the vacuum connections to the condenser, Woulff bottle,

vacuum controller with valve and vacuum pump.

• Always connect the vacuum hose (vac) to the condenser at the

highest point to minimize solvent losses during suctioning.

• Use standard laboratory vacuum hoses with an inside diameter

of 8 mm and a wall thickness of 5 mm (see accessories).

Fig. 20

Unscrew the screw (Z) fully out of the foot. (Fig. 21a)
Press the foot with the pin and the O-rings on it into the hole in
the heating bath base plate as far as the stop.
Tighten the screw (Z) by hand. (Fig.21b)

Fig. 21b

Fig. 21a

Z

• Place the heating bath on the stand of the rotation drive and

push it into the left position. (Fig. 22)

Fig. 22

Heating bath HB 10

Refer to the chapter entitled
“Commissioning“ in the heating bath
instruction manual !

• Assemble the foot of the heating bath HB10
For operation of the HB10 heating bath in combination with the
RV8, the foot plate supplied must be fitted to the heating bath.
Caution: It is not permitted to use HB10 heating bath fitted
with a foot plate in combination with any other type of rotary
evaporator.

CAUTION