• Place the RV 10.8001 condenser seal (T) in the condenser

receptacle and fit the glassware to the device according to the
assembly instructions. (Fig. 17, 18 and 19)

Mount the condenser

• Fix the receiving flask and hose connectors as shown. See also

operating manual of glassware. (Fig.19)

Fig.18

Fig.19

W

U

T

Fig.17

• Place the cap nut into the condenser, and then put the spring

ring to the flange of the condenser (U).

• Place the condenser on the rotary drive (W) and tighten the cap

nut by hand. (Fig.18)

Removing the condenser
• Use the ring spanner provided to loosen union nuts that are

tightly fitted.

• Loosen the union nut by turning anticlockwise.

Description of special condensers
(see products/accessories at www.ika.com)
• RV10.3 Vertical-intensive condenser with manifold

This vertical-intensive condenser features a double jacket design
for particularly efficient condensation.
Also available with coating (RV10.30)

• RV10.4 Dry ice condenser

Dry ice condenser for distilling low-boiling solvents. Cooling
by dry ice, no cooling water required. Maximum condensation
thanks to low temperatures.
Also available with coating (RV10.40)