Technical data, Error codes – IKA RV 8 FLEX User Manual

Page 34

Advertising
background image

34

Technical data

Operating voltage range

Vac

(100 - 240) ± 10%

Rated

voltage

Vac

100

-

240

Frequence Hz

50

/

60

Rated power of RV8 device(without heating bath)

W

75

Rated power (max.) of switching adaptor

W

90

Speed

rpm

5

-

300

Speed tolerance

Set speed: < 100 rpm

rpm

± 1

Set speed: ≥ 100 rpm

%

± 1

Speed

display

digital

Dimensions of visible display area (W x H)

mm

37 x 18

Display

7

segments

display

Smooth

start

yes

Lift

manual

Stroke

mm

120

Head

angle

adjustable

-

45°

Cooling

surface

cm

2

1500

Cooling water flow rate min.

l/h

30

Cooling water flow rate max.

l/h

100

Cooling water pressure max.

bar

1

Perm.

On-time

%

100

Perm. ambient temperature

°C

5 - 40

Perm.

relative

humidity

%

80

Protection acc. to DIN EN 60529

IP 20

Protection

class

I

Contamination

level 2

Weight (no glassware; no heating bath)

kg

12.5

Dimensions (W x D x H)

mm

595 x 390 x 615

Operation at a terrestrial altitude

m

max. 2000

Subject to technical changes!

Error Codes

Any malfunctions during operation will be identified by an error message on the display.
Once a serious error message has been displayed, the lift moves to the top end position and the device can no longer be operated.

Proceed as follows in such cases:
- switch off device using the main switch,
- carry out corrective measures,
- restart device.

Error code

Cause

Effect

Correction

E03

Temperature inside the device
is too high

Room temperature > 40°C

- Switch off the device
- Let the device cool down
- Check the location

(see Technical data)

E04

Target speed is not reached
Motor blocked

Load too large
Cable break

- Reduce the volume in the evaporator

E09

Flash memory storage error

Transmission error

- Turn the unit off and on again

If the actions described fail to resolve the fault or another error code is displayed, then take one of the following steps:
- contact the service department,
- send the device for repair, including a short description of the fault.

Advertising
This manual is related to the following products: