IKA RV 10 control FLEX User Manual

Page 42

Advertising
background image

­­43

032010

Abbreviations used:
m

=

Numbering parameter (integer)

X

=

2

Temperature heating plate

X

=

3

Safety temperature heating plate

X

=

4

Speed

X

=

60

Interval time (1-99 seconds, 1 <= m >=99)

X

=

61

Timer (1-199 minutes, 1 <= m >=199)

X

=

62

Upper lift position (OUT_SP_62 1-> drive lift up)

X

=

63

Lower lift position (OUT_SP_62 1-> drive lift down)

X

=

66

Value vacuum controller

X

=

70

Vacuum controller hysteresis

X

=

74

Tempering medium (OUT_SP_74 0=Öl, OUT_SP_74 1=Wasser)

NAMUR Commands

Function

IN_NAME

Request designation

IN_PV_X

Read actual value

X = 2, 3, 4, 66
IN_SOFTWARE

Request software Id-number, date and version

IN_SP_X

Read target value input

X = 2, 3, 4, 60, 61, 66, 70
OUT_SP_X m

Set target value to m

X = 2, 3, 4, 60, 61, 62, 63, 66, 70, 74
RESET

Switch to normal operation

START_X

Switch on appliance (remote) function

X = 2, 4, 60, 66

STOP_X

witch off appliance function

X = 2, 4, 60, 66

PC 1.1 Cable
This cable is required to connect the 9-pin connector to a PC. (fig. 24).