1 defining the lower limit, 2 deleting the lower limit, 3 defining the upper limit – Eppendorf TransferMan4 r User Manual
Page 62: Defining the lower limit, Deleting the lower limit, Defining the upper limit, Upper limit

62
Operation
TransferMan
®
4r
English (EN)
6.9.1
Defining the lower limit
Prerequisites
• An application has been selected.
1. Position the capillary a little way above the slide.
2. Press
Z-axis Limit
.
• The Z-coordinate is marked with
L
.
• The value for the limit (
Limit Z
) is displayed.
•
Z-axis Limit
is selected.
• The capillary cannot be moved any lower.
6.9.2
Deleting the lower limit
1. Press
Z-axis Limit
.
• The limit is deleted.
6.9.3
Defining the upper limit
1. In the
Installation
menu, select the
Upper limit
submenu.
The
Upper Limit
window appears.
2. Move the capillary to the top position.
3. Save the position with
Set
.
• The Z-coordinate is marked with
U
.
• The value for the limit (
Lim
) is displayed.
• The capillary cannot be moved any higher.
Upper Limit
Menu / Installation / Upper Limit
Back
Set
Clear
X
Y
Z
Lim
=
=
=
0
0
2671
2671
µm
µm
µm U
µm