7 troubleshooting, 1 general errors, 1 motor module – Eppendorf TransferMan4 r User Manual
Page 79: 2 capillary, 3 control board and display, Troubleshooting 7.1, General errors 7.1.1, Motor module, Capillary, Control board and display
79
Troubleshooting
TransferMan
®
4r
English (EN)
7
Troubleshooting
7.1
General errors
7.1.1
Motor module
7.1.2
Capillary
7.1.3
Control board and display
Symptom/message
Cause
Remedy
Motor axes move in 
the wrong direction or 
do not match joystick 
movement.
• Motor module mounted
incorrectly.
• Incorrect installation
parameters entered.
• Direction of movement of
axis inverted.
Compare the installation 
parameters with the 
structure of the motor 
module.
Check the orientation and 
structure of the modules.
Check the connections of the 
modules at the control 
board.
Cancel the inversion of the 
axis.
Symptom/message
Cause
Remedy
Capillary moves too 
fast or too slow.
• Radius of working range
incorrectly set.
Set the radius using the 
selection dial or in the 
Speed
menu.
• Incorrect acceleration factor.
In the
Installation
menu, set
the value for the
Dyn-factor
parameter.
Capillary only moves to 
the side or vertically.
• Y-axis is disabled.
The
Y off
function disabled.
Capillary does not 
move down far 
enough.
• The
Z-axis Limit
function is
active.
• Capillary is incorrectly
adjusted.
Deactivate the
Z-axis Limit
function.
Readjust the capillary.
Symptom/message
Cause
Remedy
Device does not 
respond to keystroke 
when 
Home
function is
active.
• Function is active.
Press the
home
key again.
The capillary moves down.
Press the
Back manual
softkey.
Move the joystick.