11 speed function, 1 speed menu and parameters, 2 setting the parameter for speed – Eppendorf TransferMan4 r User Manual
Page 64: Speed function 6.11.1, Speed menu and parameters, Setting the parameter for speed

64
Operation
TransferMan
®
4r
English (EN)
6.11
Speed
function
In the
Speed
menu you can set the size of the working range and the speed at which
certain positions are approached or movements executed.
6.11.1
Speed
menu and parameters
Abb. 6-1:Menu –
Fig. 6-1:
Menu –
Speed
6.11.2 Setting the parameter for
Speed
1. Select the desired parameter.
2. Set the value for the parameter.
3. Close the menu.
Parameter
Value
Range of values Increment Standard
Coarse
Set size in μm
5 – 12500
5
6000
Fine
Set size in μm
5 – 2000
5
250
X-fine
Set size in μm
0 – 600
1
80
Step inj. speed
Set speed in μm per second
5 – 10000
5
300
Position speed
Set speed in μm per second
5 – 10000
5
1500
Home speed
Set speed in μm per second
5 – 10000
5
7500
If a value of 0 is set for the parameter
X-fine
, then the option of changing
between the
Fine
and
X-fine
working ranges is deactivated.