background image

15

4 Maintenance

The

in situ

Adapter does not require regular maintenance.

Contamination can be removed using water or a mild laboratory cleaning
agent. The covering hood should be wiped using a moist cloth only in
order to prevent water or other liquids from entering the lateral ventilation
slits of the double chamber.

Do not allow any parts of the

in situ

Adapter to come into contact with

organic solvents or aggressive chemical. Do not clean the transparent
plastic parts using alcohol or acetone.

5 Technical data

Type of instrument:

in situ

Adapter for Mastercycler

or Mastercycler

gradient

Sample capacity:

Max. 4 glass slides of 76 x 26 mm
(3 x 1 inch)

Temperature range:

4 to 99 °C

Temperature distribution

40 °C: ± 0,6 °C

across the adapter:

72 °C: ± 0,6 °C
95 °C: ± 1,0 °C

Temperature-control speed:

Heating rate

1,0 °C/s,

measured on the adapter
Cooling rate

0,7 °C/s,

measured on the adapter

Dimensions of the adapter:

Metal carrier:

Width: 112 mm
Depth:

79 mm

Height:

9 mm

Covering hood: Width: 148 mm

Depth: 105 mm
Height: 31 mm

Weight:

Metal carrier: approx. 60 g

Conditions of use:

0 to 105 °C,
25 to 100 % relative humidity

Storage conditions:

–20 to +70 °C,
25 to 85 % relative humidity