MF60

– 1.04

Furnace Overheats (continued)
The Heat
lamp/s remain
on when the
controller is off

► The SSR has

failed

‘ON’

►

Check for an accidental wiring
fault which
could have overloaded the
SSR

A temperature
that is higher
than the
set-point is
shown during
the test

► No fault. Burning bitumen

causes the temperature to
increase above the set point.
This is normal.

Chimney Emits Smoke
Some light
smoke is
emitted
in the middle of
the test

► No fault

Dense smoke is
emitted

► The after-burner elements

or control may have failed

►

Investigate or contact
Carbolite

► A fuse may have blown

►

Check the Supply fuses

Warning Lamp & Door Faults
Red Electric
Shock Risk
lamp
lights when
door is opened

► Failed cut-off relay SR1

or SR2

►

DO NOT USE

until relay

replaced

► Fuse for left side of

chamber blown

►

Check and replace fuse

Red Electric
Shock Risk
lamp
does not light
when tested

►

Faulty lamp

►

DO NOT USE until lamp
replaced

► Faulty

relay SR3 or SR4

►

DO NOT USE

until relay

replaced

► Faulty momentary switch

►

DO NOT USE until switch
replaced

►

Faulty door switch

►

DO NOT USE until door
switch
replaced

Door does not
unlock

► Furnace may have been

turned off during a test or
there may have been a
power failure during a test

►

Enter START PROCESS then
cancel the test.
THE RELEASE CODE ‘728’ is
REQUIRED

► Failure of interface board

► Loose connection lead

►

Check all leads