Carbolite ABA 7/35B User Manual

Page 68

Advertising
background image

68

MF60

– 1.04

Warning Lamp & Door Faults (continued)

Door does not
lock at start of
test

► The key has been used

and left in unlocked
position

►

Set the manual override to the
locked position

► Door switch failure

►

► Interface board failure

► Loose connection lead

►

Check all leads

Balance Errors
Weight
readings not
stable

► Balance not properly

installed.

►

Re-install balance

► Balance pan extension not

free to move

►

Check for obstructions such
as
aggregate in the balance pan
extension holes

► Loose wires obstructing

balance pan

►

Check control compartment

► Unit not on firm base

►

Reposition ABA or adjust
stand

► Fan motor support bracket

has failed

►

Check/adjust motor support
bracket

► Fan failure causing vibration

►

Check/replace motor or
impellor

Fan Problems
Fan not
switching off

► Faulty relay FS in fan circuit

►

Do not use until the relay is
replaced

► Faulty Control Board

►

Check diagnostics screen.
Ref. Error! Reference
source not found.

Fan not
switching on

► Main chamber temperature

less than 100°C

►

Check heater is on

► Door is open

►

Close door

► Test not in progress

►

Run test

► Fan motor failure

►

Replace motor

► Faulty relay FS

Replace relay

► Faulty Control Board

► Loose connection lead

►

Check all leads

Test Problems
Early
termination of
test

► Test parameters of trigger

weight set too high

►

Adjust to suitable setting

► Test parameter of constant

weight delay set too low or
zero

►

Adjust to correct setting and
save;
delete any old setting

► Balance unstable

►

See balance errors

Printer Problems
Printer does not
print

► Printer is out of paper

►

Add more printer paper

► Faulty printer supply switch

►

Replace switch

► Faulty printer

►

Replace printer

► Loose lead

►

Check connector at back of
printer

► Wrong settings in printer

►

Run ‘Printer Test’ 12