20

Calibrating (MR Hei-Tec)

With the calibrating function, you can change the temperature of the medium measured using the
temperature sensor Pt 1000 by a calibration value of -5 °C to +5 °C.

Stirring function and heating function

The stirring function and the heating function may not be active.


1. Connect temperature sensor (see chapter "Installation" – "Accessories (optional)" - „Temperature sensor

Pt 1000").

2. Calibrate temperature.

Example: 31 °C




Press and hold
Hei Guide

Press the Hei Guide
rotary-push button
and hold button
during the entire
setting procedure.


Press Stirring
button

Now press the
Stirring button as
well.

Increase/decrease
calibration value

To increase or
decrease the
calibration value,
press the Stirring
button repeatedly
until you reach the
desired calibration
value.