Error codes

NOTICE

Error code displayed on the device

1. Perform a reset.
2. Implement the measures suggested in the table below.

MR Hei-Standard

Error code

Cause

Consequence

Measure

LED I/O flashes
one time

-

Heating function off.  Please contact our

Technical Service.

LED I/O flashes
two times

The actual temperature
exceeds the set
temperature on the sensor of
the temperature sensor by
more than 25 °C:

a. Temperature control

defective.

b. Exothermic reaction.
c. Test setup not optimal.

 Check test setup.
 Please contact our

Technical Service.

LED I/O flashes
three times

a. Speed control knob

defective.

b. Temperature control knob

defective.

c. Heating button defective.

Heating function off.  Please contact our

Technical Service.

LED I/O flashes
four times

Motor defective.

Heating function off.
Stirring function off.

LED I/O flashes
five times

a. Sensor break.
b. Temperature sensor is

connected or disconnected
whilst the heating function
is switched on.

(IMPORTANT: The tempera-
ture sensor must only be
connected or removed when
the heating function is
switched off)

Heating function off.  Replace temperature

sensor (only in case of
“a”.).

 Switch off the device.

 Connect temperature

sensor (see chapter
"Installation

").

 Switch on the device.

MR Hei-Tec

Error

code

Mode

Cause

Consequence

Measure

„E11“,
„E12“,
„E13“,
„E14“

-

Heating function off.  Please contact our

Technical Service.