3 recorder outputs, 2 interface settings at the ecotec e3000, 1 recorder outputs – INFICON Ecotec E3000 Multigas-Sniffer-Leak Detector Interface description User Manual

Page 10: Recorder outputs, Interface settings at the ecotec e3000

Advertising
background image

10

2 I/O Port

In

te

rface D

e

scr

ipti

o

n

Ecotec E300

0,

k

ins

22e1

-j

, 1401

Technical Data relay
outputs

Relay output

Active = Normally open contact (NO) closed

Trigger

If the preset trigger level is exceeded, relay contact NO will close and the relay contact NC
will open respectively.

Ready

As long as the Ecotec E3000 is ready for measurements, the relay contact NO is closed and
the relay contact NC is open respectively.

2.1.3

Recorder outputs

Analogue output

0 … 10 V

max. 1 mA

2.2

Interface settings at the Ecotec E3000

At “Settings > Interfaces” you make the settings for the interfaces.

2.2.1

Recorder outputs

For the recorder output the “Recorder Output Scale” and the “Recorder Output Gas” can
be selected.

The “Recorder Output Scale” can either be linear or logarithmic. The output signal is
provided through channel 1 (pin 1 of the I/O port).

Which of the 4 gases is indicated here is defined through the menu line “Recorder Output
Gas”.

Recorder output scale

Example:

The trigger level of the corresponding gas has been set to 3 g/a.

Material damage if load is high.

Wrong input voltage or current can destroy the unit.

Pay attention to the technical data of the relay outputs: Maximum load rating is
60 V DC / 25 V AC and 1 A per relay (resistive load).

LIN

The voltage range spans from 0 V to 10 V.

The leak rate which is output is normalized with respect to the trigger
level for the corresponding gas. The trigger level will correspond to an
output of 1 V. Higher and smaller leak rates may be output as a voltage
deviating by a corresponding factor with reference to the trigger level.

leak rate = 0.3 g/a

 U

analog

= 0.1 V

leak rate = 3 g/a

 U

analog

= 1.0 V

leak rate = 10 g/a

 U

analog

= 3.3 V

Voltages over 1 V indicate that the trigger level has been exceeded.