3 RS-232 interface

In

te

rface D

e

scr

ipti

o

n

Ecotec E300

0,

k

ins

22e1

-j

, 1401

*cal:status? (CR)

Requests the status of the calibration.

LEAK STABLE, CONFIRM (CR) Status: waiting for acknowledgement that signal from test leak is

stable.

*cal:read? (CR)

Requests leak rate signal reading.

8.2638e-14 (CR)

Signal from test leak is 8.2638e-14.

..........

This request should be repeated several times to determine whether the reading has stabilized.

*cal:quit (CR)

Starts the test leak measurement.

OK (CR)

Ecotec E3000 has received start signal.

*cal:status? (CR)

Requests the status of the calibration.

WAIT (CR)

Status: measuring test leak

..........

*cal:status? (CR)

Requests the status of the calibration.

AIR STABLE, CONFIRM (CR)

Status: waiting for background measurement, waiting for signal to
be confirmed to be stable.

*cal:read? (CR)

Requests leak rate signal reading.

3.0513e-15 (CR)

Signal from background is 3.0513e-15.

..........

This request should be repeated several times to determine whether the reading has stabilized.

*cal:quit (CR)

Starts the background measurement.

OK (CR)

Ecotec E3000s has received start signal.

*cal:status? (CR)

Requests the status of the calibration.

WAIT (CR)

Status: measuring background

..........

*cal:status? (CR)

Requests the status of the calibration.

CAL FINISHED, CONFIRM (CR) Status: calibration finished without errors, results are available.

*cal:factor:old? (CR)

Requests data for old calibration factor.

1.95 (CR)

Old calibration factor was 1.95.

*cal:factor:new? (CR)

Requests data for new calibration factor.

2.05 (CR)

new calibration factor is 2.05.

*cal:pos:old? (CR)

Requests data for old mass position deviation.

0.05 (CR)

Old mass position deviation was 0.05.

*cal:pos:new? (CR)

Requests data for new mass position deviation.

0.10 (CR)

New mass position deviation is 0.10.

*cal:flow:old? (CR)

Requests data for old flow at calibration.

176 (CR)

Old flow was 176 (sccm).

*cal:flow:new? (CR)

Requests data for new flow at calibration.

187 (CR)

New flow is 187 (sccm).

*cal:quit (CR)

Save results of new calibration.

OK (CR)

Ecotec E3000 has received request to save new values.

*cal:status? (CR)

Requests the status of the calibration.

WAIT (CR)

Status: wait, data is being saved, new offset is determined.

*status? (CR)

Requests the status of the Ecotec E3000.

Table 14: Commands and responses external calibration (cont.)

Command sent to

Ecotec E3000

Response received from

Ecotec E3000

Meaning