3 RS-232 interface

19

In

te

rface D

e

scr

ipti

o

n

Ecotec E300

0,

k

ins

22e1

-j

, 1401

:SELect

gas select (Auto, Manu, Hold)

R/S

:SELTime

Gas select time (5, 10, 15, 20)

R/S

:SENSitivity

sensitivity check (ON, OFF)

R/S

:SNIFFer

Audio sniffer (off, TRIGger SEARch)

(for compatibility TRIGger

 on)

R/S

:SPEAker

audio speaker internal (ON, OFF)

R/S

:UNIT

pressure unit (Torr, mbar, atm, Pa)

R/S

:VOLMin

minimum volume (0 … 15)

R/S

:VOLume

volume (0 … 15)

R/S

:WAKEup

:1 … :7

:MON, :TUE…

WakeUp (hh:mm, 00:00

 disable)

1-Monday, 2-Tuesday, …

R/S

:WARNreminder

warning reminder (ON, OFF)

R/S

:ZERO

ZERO (ON, OFF)

sniffer line and main unit affected

:ZEROSniff

ZERO sniffer line (ON, OFF)

R/S

:ZEROMain

ZERO via main unit (ON, OFF)

R/S

:ZEROTime

ZERO time (1.0 … 9.9)

R/S

*GAS:<no> 1 … 4

gas number (1 … 4)

:CALFAC

calibration factor

R

:CALintern

Internal calibration

enable: on, 1, ENAble

disable: off, 0, DISable

(depends on selected mass)

R/S

:CONFig

Configuration selected mass position 1, 2, … 5

(depends on masses and factor)

R/S

:CORRfac

Correction factor (The correction factor is calculated
from the other settings)

R

:EQUIname

Equivalent gas name

(Equivalent gas only possible, if He/H2 mass 2/4 is
selected. If an equivalent gas name is set, the viscosity
is loaded from the RomLib.

R/S

:FACTOR

Factor to the mass positions (5)

(example: 0.72, 0.46, 0.12, 0.0, .0.)

R

:GASpress

Equivalent gas pressure [bar]

R/S

:HEpress

Pressure (He / H2) [bar]

R/S

:LASTcal

last calibration (Date, Time, Internal / External)

R

:LIMIT

Lower display limit (1, 2, 5, 10, 20, 50, 100)

:LRlimit

Lower display limit (leak rate)

R

:MASS

mass position [amu] ± deviation [amu]

R

:MASSES

possible mass positions (5)

(example: 69, 83, 51, 0, 0) (R600a IGS on position 5)

Notice: Changing this setting will clear the calibration
data!

R/S

Table 11: List of commands, ASCII (cont.)

Command

word1

Command

word2

Command

word3

Statement

Read/Set

Advertising