Hardware, 1 specifications, 2 connections – INFICON STM-1 Thin Film Deposition Monitor User Manual

Page 3: 3 dipswitch settings, 1 . hardware

Advertising
background image

1. Hardware


1.1 Specifications:

Sensor Inputs: 1
Sensor Frequency: 1.95MHZ - 10.05MHZ
Frequency Accuracy (Absolute): <2ppm
Frequency Stability (Relative Accuracy) Nominal: .03HZ
Measurement Period: 100ms (Ten measurements per second, regardless of frequency).

Thickness Resolution:

.02 A


Rate (Raw) Resolution: .02 A/sec

Communication Electrical: RS232 or RS485 (dip-switch selectable), 9600 or 115200 BAUD (dip-switch selectable)
Communication Protocol: SMDP (Sycon Multi Drop Protocol)

1.2 Connections:


Electrical Connections at board (DB9 female connector):
Pin# Description
1

Analog output ground. NOT ISOLATED.

2

RS232 TxD/RS485 inverting (-)

3

RS232 RxD/RS485 non-inverting (+)

4 No

connection

5 Communication

signal

gnd.

6

Analog output, 0 to +10V, 12 bits, nominal. 30 mA current source.

7 No

connection

8 No

connection

9

Alternate power (+5V) input if J2 is not used, 500 mA supply required.



RS232 PC to STM1 communication cable:
DB9 male pin# (plug into STM1)

Connect to

DB9 FEMALE PIN# (plug into PC)

2 (STM1 TxD)

2 (PC RxD)

3 (STM1 RxD)

3 (PC TxD)

5 (STM1 GND)

5 (PC GND)

1.3 Dipswitch Settings:

On, or 1, is when the dipswitch is in the up position (top position when switch is viewed with switch labeled 1 on the left and
switch 8 on the right). Off, or 0, is when the dipswitch is in the down position.

Switch number

Controls

Description

1 Nothing

Unused

2

Endian Selection (see note 1.1)

0 = little endian (LS first)
1 = big endian (MS first), Default

3 Communication

Checksum

(See note1.2)

0 = No checksum
1 = Checksum required, Default

4

Baud Rate

0 = 9600 Baud
1 = 115,200 Baud, Default

3